1. Giới thiệu
Thank you for choosing the FORA Painless Design Adjustable Lancing Device. This device is designed to assist in obtaining blood samples for glucose monitoring with enhanced comfort and safety. Please read this manual thoroughly before using the device to ensure proper operation and to understand all safety precautions.
Các tính năng chính:
- Cài đặt độ sâu có thể điều chỉnh: Offers 6 dial-in-click depth settings for personalized comfort based on skin type.
- Safe Lancet Ejection: Features a built-in ejector mechanism for hygienic removal of used lancets without direct contact.
- Kiểm tra trang web thay thế: Includes a clear end cap for testing on alternative sites such as the forearm or palm.
- Khả năng tương thích: Compatible with most standard lancets (lancets sold separately).
2. Thông tin an toàn
Always follow basic safety precautions when using this lancing device to reduce the risk of injury or infection.
- Chỉ sử dụng cho một bệnh nhân duy nhất: Thiết bị lấy máu này chỉ dành cho một người sử dụng. Việc dùng chung thiết bị có thể dẫn đến lây truyền các mầm bệnh qua đường máu.
- Use New Lancets: Always use a new, sterile lancet for each test. Reusing lancets can cause infection and dull the tip, leading to increased pain.
- Xử lý đúng cách: Dispose of used lancets in an approved sharps container immediately after use.
- Để xa tầm tay trẻ em: Cất giữ dụng cụ chích máu và kim chích ở nơi an toàn, tránh xa tầm tay trẻ em.
- Tham khảo ý kiến chuyên gia y tế: Các tuyên bố liên quan đến thực phẩm chức năng chưa được FDA đánh giá và không nhằm mục đích chẩn đoán, điều trị, chữa khỏi hoặc ngăn ngừa bất kỳ bệnh hoặc tình trạng sức khỏe nào. Luôn luôn tham khảo ý kiến chuyên gia chăm sóc sức khỏe để được tư vấn y tế.
3. Sản phẩm hếtview
Familiarize yourself with the components of your FORA Lancing Device:

Hình 3.1: Components of the FORA Lancing Device. This diagram illustrates the main parts: the protective cap, the depth indicator for setting penetration level, the ejector mechanism for safe lancet removal, and the release button to activate the lancing action.
- Mũ lưỡi trai: The protective cover that twists off to access the lancet chamber.
- Chỉ số độ sâu: A dial with 6 settings to adjust the lancet penetration depth.
- đầu phun: A mechanism to safely remove used lancets without direct handling.
- Nút phát hành: The button pressed to activate the lancing action.
- Lancing Device Body: The main housing of the device.
4. Thiết lập
4.1 Inserting a Lancet
- Unscrew the cap from the lancing device.
- Lắp chặt kim chích máu mới, đã được khử trùng vào giá đỡ kim chích.
- Twist off the protective cap from the lancet. Retain this cap for safe disposal later.
- Replace the lancing device cap by screwing it back on.
4.2 Adjusting Penetration Depth
The lancing device features 6 adjustable depth settings. Rotate the cap to select the desired depth. A higher number indicates a deeper penetration.
- Cài đặt 1-2: For soft or thin skin.
- Cài đặt 3-4: For average skin.
- Cài đặt 5-6: For thick or calloused skin.

Hình 4.1: Steps for preparing and using the lancing device. This image demonstrates selecting the penetration depth, cocking the device, and performing a finger prick.
5. Hướng dẫn vận hành
5.1 Performing a Blood Sample (Fingerstick)
- Hãy rửa tay thật kỹ bằng xà phòng và nước ấm, sau đó lau khô hoàn toàn.
- Ensure a new lancet is inserted and the depth setting is appropriate for your skin type (refer to Section 4.2).
- Pull the cocking control back until it clicks. The device is now armed.
- Đặt chặt dụng cụ lấy máu vào cạnh đầu ngón tay của bạn.
- Nhấn nút nhả để kích hoạt kim chích máu.
- Gently massage your finger from the base towards the tip to form a blood drop. Do not squeeze excessively.
- Apply the blood drop to your glucose meter test strip as instructed by your meter's manual.
5.2 Alternative Site Testing (AST)
The FORA Lancing Device includes a clear alternate site end cap for testing on areas other than the fingertip, such as the forearm or palm. Consult your healthcare professional before performing AST, as results from alternative sites may differ from fingertip results, especially during periods of rapid glucose change.
- Remove the standard cap and replace it with the clear AST cap.
- Select a suitable depth setting (AST generally requires a deeper setting due to thicker skin).
- Pull the cocking control back until it clicks.
- Press the clear AST cap firmly against the chosen alternative site (e.g., forearm or palm) and hold for a few seconds to create a vacuum.
- Nhấn nút phát hành.
- Gently massage the area to form a blood drop.
- Apply the blood drop to your glucose meter test strip.
5.3 Safely Disposing of a Used Lancet
Proper disposal of used lancets is crucial to prevent accidental needle sticks and the spread of infection.

Hình 5.1: Steps for safe lancet disposal. This image shows how to cap the lancet and then use the ejector mechanism to remove it from the device.
- Tháo nắp thiết bị lấy máu.
- Carefully place the protective cap (removed in Section 4.1) back onto the used lancet, ensuring the needle is fully covered. Alternatively, if your lancets come with a protective disk, push the exposed tip into the disk on a hard surface.
- Slide the ejector forward to release the capped lancet into an approved sharps container.
- Thay nắp thiết bị chích máu.
6. Bảo trì
6.1 Cleaning the Lancing Device
- Clean the exterior of the lancing device regularly with a soft cloth dampđược làm sạch bằng xà phòng nhẹ và nước.
- Không được nhúng thiết bị vào nước hoặc bất kỳ dung dịch tẩy rửa nào.
- Đảm bảo thiết bị khô hoàn toàn trước khi cất giữ hoặc sử dụng lần sau.
6.2 Lưu trữ
- Store the lancing device in a clean, dry place at room temperature.
- Tránh nhiệt độ quá cao, ánh nắng trực tiếp và độ ẩm cao.
- Để thiết bị và tất cả các phụ kiện xa tầm tay trẻ em và vật nuôi.
KHAI THÁC. Xử lý sự cố
If you encounter issues with your lancing device, refer to the table below for common problems and solutions.
| Vấn đề | Nguyên nhân có thể | Giải pháp |
|---|---|---|
| Device does not cock or fire. | Device not fully cocked; lancet not inserted correctly; device malfunction. | Ensure the cocking control is pulled back until it clicks. Re-insert lancet. If problem persists, contact customer support. |
| Không đủ máuamplà. | Depth setting too shallow; insufficient blood flow; cold hands. | Increase depth setting. Warm hands before testing. Gently massage the lanced area. |
| Pain during lancing. | Depth setting too deep; reused lancet; lancing on calloused area. | Decrease depth setting. Always use a new lancet. Choose a different lancing site. |
| Lancet cannot be ejected. | Lancet stuck; ejector mechanism obstructed. | Ensure the cap is removed. Gently try to slide the ejector. If stuck, carefully remove with tweezers and dispose of safely. |
8. Thông số kỹ thuật
| Tính năng | Chi tiết |
|---|---|
| Người mẫu | Painless Design Adjustable Lancing Device |
| Cài đặt độ sâu | 6 mức điều chỉnh |
| Lancet Compatibility | Most standard lancets (28-33 gauge) |
| Kích thước | Khoảng 6.1 x 3.2 x 1 inch |
| Cân nặng | Khoảng 1.6 ounce |
| Nhà sản xuất | Công ty ForaCare Inc. |
| ASIN | B01INS8T5I |
9. Bảo hành và Hỗ trợ
For warranty information, technical support, or to purchase accessories, please contact FORA customer service or visit the official FORA webđịa điểm.
You can find more information about FORA products and support at the Cửa hàng FORA trên Amazon.