The YSI 251398Y DPD Free Chlorine Reagent Powder is designed for accurate measurement of free chlorine in water samples using a compatible colorimeter. Each pack contains 100 individual powder packs, with each pack formulated for a 5ml sample volume and capable of measuring free chlorine in the range of 0 to 3.5 mg/L.

This reagent utilizes the DPD (N,N-Diethyl-p-phenylenediamine) method, which reacts with free chlorine to produce a magenta color, the intensity of which is proportional to the chlorine concentration and measured by a colorimeter.

Thông tin an toàn

Always handle chemical reagents with care. Read the Safety Data Sheet (SDS) for this product before use. Wear appropriate personal protective equipment (PPE), including safety glasses and gloves, when handling the reagent powder. Avoid ingestion and contact with skin or eyes. In case of contact, rinse thoroughly with water and seek medical attention if irritation persists.

Thiết lập và Chuẩn bị

Before beginning any measurements, ensure you have the following:

  • YSI 251398Y DPD Free Chlorine Reagent Powder pack
  • Compatible colorimeter or spectrophotometer
  • sạch samptế bào/lọ
  • Deionized or distilled water for blanks and dilutions (if necessary)
  • Nhiều nướcample sẽ được kiểm tra
  • Bộ đếm thời gian

Các bước chuẩn bị:

  1. Ensure all glassware and sample cells are clean and free from contamination. Rinse with deionized water before use.
  2. Calibrate your colorimeter according to its specific instruction manual before use.

Hướng dẫn sử dụng

Follow these steps to accurately measure free chlorine using the YSI 251398Y DPD Free Chlorine Reagent Powder:

  1. SampBộ sưu tập: Thu thập một nguồn nước ngọtample in a clean sample cell. The required volume is 5ml. Ensure the sample is representative of the water being tested.
  2. Thêm thuốc thử: Carefully tear open one YSI 251398Y DPD Free Chlorine Reagent Powder pack. Add the entire contents of the powder pack to the 5ml water sample trong sampô le.
  3. Trộn: giới hạn sample cell and invert several times to thoroughly mix the reagent with the sample. Ensure all powder dissolves completely. A magenta color will develop if free chlorine is present.
  4. Thời gian phản ứng: Allow a reaction time of 1 minute for the color to fully develop. Do not exceed 6 minutes, as color fading may occur, leading to inaccurate results.
  5. Đo lường: Chèn các phần đã chuẩn bị vàoample cell into your calibrated colorimeter. Follow your colorimeter's instructions to initiate the measurement. The colorimeter will display the free chlorine concentration in mg/L.
  6. Kết quả ghi nhận: Record the measured free chlorine concentration.
Close-up of a silver foil reagent powder packet with 'VARIO.LIT Chlorine FREE-DPD' printed on it.

Image: A single YSI DPD Free Chlorine Reagent Powder packet. This silver foil packet contains the pre-measured reagent powder for a single test. The text "VARIO.LIT Chlorine FREE-DPD" is visible on the packet, indicating its contents and purpose.

Bảo trì và lưu trữ

Proper storage and handling are crucial for maintaining the integrity and effectiveness of the reagent powder:

  • Kho: Store the reagent powder packs in their original packaging in a cool, dry place, away from direct sunlight and extreme temperatures. Do not store near strong acids or bases.
  • Hạn sử dụng: Refer to the expiration date printed on the packaging. Do not use expired reagents, as they may yield inaccurate results.
  • Xử lý: Dispose of used reagent and sample solutions according to local environmental regulations. Consult your local authorities for proper chemical waste disposal procedures.
  • Độ sạch sẽ: Always ensure sample cells and equipment are thoroughly cleaned after each use to prevent cross-contamination.

Xử lý sự cố

If you encounter issues during testing, consider the following common problems and solutions:

Vấn đềNguyên nhân có thểGiải pháp
No color developmentNo free chlorine present; Expired reagent; Incorrect sampkhối lượng le.Xác minh sample contains chlorine; Check reagent expiration date; Ensure 5ml sampkhối lượng le.
Fading color or inaccurate readingReaction time exceeded; Contaminated sample cell; Turbidity in samplà.Adhere strictly to 1-minute reaction time; Clean sample cells thoroughly; Filter turbid samples nếu cần thiết.
Kết quả không nhất quánImproper mixing; Temperature variations; Colorimeter not calibrated.Ensure complete dissolution of powder; Perform tests at consistent temperatures; Recalibrate colorimeter.

Thông số kỹ thuật

  • Tên sản phẩm: YSI DPD Free Chlorine Reagent Powder
  • Số hiệu mẫu: 251398Y
  • Reagent Type: DPD (N,N-Diethyl-p-phenylenediamine)
  • Analyte: Clo tự do (Cl2)
  • Sample Khối lượng: 5ml
  • Phạm vi đo lường: 0 đến 3.5 mg / L (ppm)
  • Bao bì: Powder packs, 100 per pack
  • Nhà sản xuất: YSI
  • ASIN: B00CMX0PDK
  • Kích thước gói hàng: 8.1 x 6.9 x 1.2 inch
  • Trọng lượng sản phẩm: 3.2 ounce

Bảo hành và Hỗ trợ

For information regarding product warranty, technical support, or to purchase additional reagents, please contact YSI customer service or visit the official YSI website. Refer to your purchase documentation for specific warranty terms and conditions.

YSI Contact Information:

  • Webđịa điểm: www.ysi.com (Please note: This is a placeholder link, refer to official YSI documentation for exact contact details.)

Tài liệu liên quan - 251398Y

Trướcview Sổ tay hướng dẫn sử dụng máy đo độ dẫn điện YSI Model 32
Sách hướng dẫn sử dụng Máy đo độ dẫn điện YSI Model 32, trình bày chi tiết mô tả chung, thông số kỹ thuật, quy trình vận hành, bảo trì và hiệu chuẩn.
Trướcview Hướng dẫn nhanh về hiệu chuẩn và vận hành máy đo YSI Pro20/Pro20i:
Hướng dẫn ngắn gọn về hiệu chuẩn và vận hành máy đo oxy hòa tan YSI Pro20/Pro20i, bao gồm lắp đặt màng lọc, thiết lập cảm biến, hiệu chuẩn áp kế và DO, bù độ mặn và thực hiện phép đo. Bao gồm các bước vận hành thiết yếu và thông tin liên hệ.
Trướcview Bảng dữ liệu an toàn (SDS) của dung dịch đệm YSI 3821 pH 4.00
Bảng dữ liệu an toàn toàn diện (SDS) cho dung dịch đệm YSI 3821 pH 4.00, nêu chi tiết thông tin nhận dạng sản phẩm, mối nguy hiểm, thành phần, sơ cứu, xử lý, lưu trữ và quy định.
Trướcview Hướng dẫn sử dụng thiết bị đo nồng độ oxy hòa tan và nhiệt độ YSI EcoSense DO200A/DO200M
Sách hướng dẫn sử dụng này cung cấp hướng dẫn toàn diện về vận hành, lắp đặt, bảo trì và khắc phục sự cố cho thiết bị đo nồng độ oxy hòa tan và nhiệt độ YSI EcoSense DO200A và DO200M. Sách bao gồm thông tin chi tiết về thông số kỹ thuật, phụ kiện và thông tin bảo hành.
Trướcview Hướng dẫn vận hành và bảo trì máy phát vệ tinh YSIT-1
Sách hướng dẫn sử dụng toàn diện cho bộ phát tín hiệu vệ tinh YSI Model YSIT-1, bao gồm vận hành, bảo trì, thiết lập phần cứng và tham khảo lệnh nối tiếp. Chi tiết bao gồm quản lý thời gian, lưu trữ dữ liệu, định dạng dữ liệu GOES, điều khiển bộ phát RF và đồng bộ hóa GPS.
Trướcview YSI 9300 Photometer Operations Manual
Comprehensive operations manual for the YSI 9300 direct-reading photometer, covering its principles, operation, maintenance, and troubleshooting for accurate water quality analysis.